Download or browse on-line these Service And Technical Reference Manual for Respironics BiPAP Synchrony Medical Equipment.
This manual for Respironics BiPAP Synchrony, given in the PDF format, is available for free online viewing and download without logging on. The guide contains 155 pages, and the size of the file at download is . The document type is Service And Technical Reference Manual.
In case you failed to obtain relevant information in this document, please, look through related operating manuals and user instructions for Respironics BiPAP Synchrony. Just click one of the links below to go to the selected manual:
[Page 1] Respironics BiPAP Synchrony
B B iPAP® iPAP® Synchrony Synchrony Service Service & Technical & Technical Information Information .
[Page 2] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REVISION HISTORY - PAGE 11032450, REV. 02 REVISION HISTORY SECTION NAME (NO.) REV. # DATE DESCRIPTION Copyright/Warranty (1032405) 00 Initial Revision Revision History (1032450) 02 Initial Revision .
[Page 3] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032450, REV. 02REVISION HISTORY - PAGE 2 This page intentionally blank. .
[Page 4] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION COPYRIGHT/WARRANTY - PAGE 11032405, REV. 00 © 2006, Respironics, Inc. and affiliates. All rights reserved. .
[Page 5] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032405, REV. 00COPYRIGHT/WARRANTY - PAGE 2 WARRANTY Respironics, Inc. warrants that the system shall be free from defects of workmanship and materials and will perform in accordance with the produc.
[Page 6] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TABLE OF CONTENTS - PAGE 11032406, REV. 01 INTRODUCTION BIPAP SYNCHRONY VENTILATORY SUPPORT SYSTEM OVERVIEW . 1 SERVICE NOTICE .
[Page 7] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032406, REV. 01TABLE OF CONTENTS - PAGE 2 SLEEVE VALVE . 4 VALVE CONTROLLER .
[Page 8] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TABLE OF CONTENTS - PAGE 31032406, REV. 01 PATIENT ALARMS . 7 POWER ALARMS .
[Page 9] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032406, REV. 01TABLE OF CONTENTS - PAGE 4 TESTING TESTING PROCESS . 1 BIPAP SYNCHRONY SYSTEM RU.
[Page 10] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION INTRODUCTION - PAGE 11032407, REV. 00 INTRODUCTION BIPAP SYNCHRONY VENTILATORY SUPPORT SYSTEM OVERVIEW The BiPAP Synchrony, shown in Figure A, is a low-pressure, electronically driven ventilator sys.
[Page 11] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032407, REV. 00INTRODUCTION - PAGE 2 The S/T mode ensures that patients receive a minimum number of breaths per minute based on the Rate setting. If the patient fails to initiate an inspiration w.
[Page 12] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION INTRODUCTION - PAGE 31032407, REV. 00 TRAINING Respironics offers service training for BiPAP and CPAP systems. For more information, Email Respironics at: [email protected] Visit R.
[Page 13] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032407, REV. 00INTRODUCTION - PAGE 4 This page intentionally left blank. .
[Page 14] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION WARNINGS & CAUTIONS - PAGE 11032408, REV. 01 WARNINGS, CAUTIONS, AND NOTES CHAPTER OVERVIEW Warnings, Cautions, and Notes are used throughout this manual to identify possible safety hazards, con.
[Page 15] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032408, REV. 01WARNINGS & CAUTIONS - PAGE 2 CAUTIONS NOTES CAUTIONS • Electronic components used in this device are subject to damage from static electricity. Repairs made to this device m.
[Page 16] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION SPECIFICATIONS & SYSTEM FEATURES - PAGE 11032409, REV. 00 SPECIFICATIONS, & SYSTEM FEATURES OVERVIEW This chapter describes the specifications and system features for the BiPAP Synchrony Ven.
[Page 17] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032409, REV. 00SPECIFICATIONS & SYSTEM FEATURES - PAGE 2 SPECIFICATIONS ENVIRONMENTAL Operating Temperature 5 °C to 35 °C Storage Temperature -20 °C to 60 °C Humidity 15 - 95% non-condensin.
[Page 18] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION SPECIFICATIONS & SYSTEM FEATURES - PAGE 31032409, REV. 00 FUNCTIONS Modes Continuous Positive Airway Pressure (CPAP) Spontaneous (S) Timed (T) Spontaneous/Times (S/T) Pressure Control (PC) PRESS.
[Page 19] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032409, REV. 00SPECIFICATIONS & SYSTEM FEATURES - PAGE 4 SYSTEM FEATURES SYSTEM FEATURES AC Power Connector AC power supply connection. Air Outlet Port Flexible tubing/patient circuit connectio.
[Page 20] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION SPECIFICATIONS & SYSTEM FEATURES - PAGE 51032409, REV. 00 RAMP FEATURE BiPAP Synchrony is equipped with a linear ramp function. The ramp feature will reduce the pressure and then gradually incr.
[Page 21] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032409, REV. 00SPECIFICATIONS & SYSTEM FEATURES - PAGE 6 A maximum IPAP time of 3.0 seconds acts as a safety mechanism to limit the time spent at the IPAP level during spontaneous breathing in .
[Page 22] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION SPECIFICATIONS & SYSTEM FEATURES - PAGE 71032409, REV. 00 SYMBOLS AND MEANINGS CONT. Electrostatic Discharge Ordinary Equipment UL Recognized for Canada and the United States TUV Safety Standard.
[Page 23] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032409, REV. 00SPECIFICATIONS & SYSTEM FEATURES - PAGE 8 This page intentionally left blank. .
[Page 24] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION THEORY OF OPERATION - PAGE 11032410, REV. 00 THEORY OF OPERATION OVERVIEW This chapter describes the operation of the BiPAP Synchrony Printed Circuit Assembly (PCA). Operation of the various circuit.
[Page 25] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032410, REV. 00THEORY OF OPERATION - PAGE 2 JTAG interface for boundry scans and a power management controller for sleep and wakeup operations. The core and I/O ports of the microcontroller use 3.3.
[Page 26] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION THEORY OF OPERATION - PAGE 31032410, REV. 00 ANALOG DEVICES PRESSURE SENSORS The main control board uses two sensors to monitor pressure amplitudes, MT4 or MT5 and MT3 or MT1. MT4 or MT5 is connect.
[Page 27] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032410, REV. 00THEORY OF OPERATION - PAGE 4 S LEEVE VALVE The sleeve valve of the BiPAP Synchrony controls with good response time and stability. The valve is a linear DC motor with a slitted slee.
[Page 28] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION SETUP - PAGE 11032411, REV. 00 SYSTEM SETUP PROCEDURES OVERVIEW This section provides an overview of the system setup, including detailed instructions on how to access the Therapy settings. F IGURE .
[Page 29] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032411, REV. 00SETUP - PAGE 2 SYSTEM SETUP Plug the external power supply into the connector in the back of the BiPAP Synchrony. The BiPAP Synchrony will sound a confirmation alarm, and the contr.
[Page 30] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION SETUP - PAGE 31032411, REV. 00 displays if the access level is set to Provider mode. The monitoring screen also displays the actual measured pressure and the Flex icon if Flex is enabled. When in t.
[Page 31] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032411, REV. 00SETUP - PAGE 4 2. If CPAP is chosen, the CPAP setting screen will appear next. Increase or decrease the CPAP pressure by pressing the Heat and Ramp buttons until the desired pressur.
[Page 32] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION SETUP - PAGE 51032411, REV. 00 4. The IPAP max screen will appear next. Increase or decrease the IPAP max pressure by pressing the Heat or Ramp buttons until the correct pressure appears. You can.
[Page 33] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032411, REV. 00SETUP - PAGE 6 7. The EPAP screen will appear next. Increase or decrease the EPAP pressure by pressing the Heat or Ramp buttons until the correct pressure appears. You can adjust t.
[Page 34] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION SETUP - PAGE 71032411, REV. 00 10. When in S mode the next screen will be the Flex Setting screen. (All other modes should proceed to Step 11.) This screen allows you to adjust the level of air .
[Page 35] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032411, REV. 00SETUP - PAGE 8 13. The Rise Time is the next screen that will appear. In S mode, this screen will only display if the Flex setting is equal to zero. Rise Time is the time it takes.
[Page 36] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION SETUP - PAGE 91032411, REV. 00 15. The next screen is the Patient Disconnect Alarm Setting. This setting enables or disables the audible alert (a beeping sound) when a large, continuous air leak (su.
[Page 37] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032411, REV. 00SETUP - PAGE 10 17. When in PC, S, S/T, and T mode and AVAPS is enabled, the next screen to appear is the Low Tidal Volume Alarm Setpoint. (For CPAP mode proceed to step 18.) This.
[Page 38] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION SETUP - PAGE 111032411, REV. 00 20. The next and final screen is the Access Level Setting screen. This screen allows you to select Provider mode or User mode access. Press the Heat or Ramp buttons .
[Page 39] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032411, REV. 00SETUP - PAGE 12 CIRCUIT CONFIGURATIONS The BiPAP Synchrony is intended for use with Respironics-approved patient circuits. Typical components are: • Bacteria Filter (Optional) �.
[Page 40] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION SETUP - PAGE 131032411, REV. 00 SOFTWARE • Respironics Encore Pro Data Management software for reading compliance data. DC POWER ACCESSORIES The BiPAP Synchrony can be used in a stationary recreat.
[Page 41] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032411, REV. 00SETUP - PAGE 14 This page intentionally left blank. .
[Page 42] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION ROUTINE MAINTENANCE - PAGE 11032412, REV. 00 ROUTINE MAINTENANCE CHAPTER OVERVIEW This chapter provides guidelines and illustrates the cleaning and maintenance procedures for the BiPAP Synchrony sys.
[Page 43] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032412, REV. 00ROUTINE MAINTENANCE - PAGE 2 F IGURE A BIPAP SYNCHRONY SYSTEM WITH FILTERS AND CAP 3. Remove the pollen filter by gently pulling the edges of the filter. Rinse the filter in a stead.
[Page 44] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032412, REV. 00ROUTINE MAINTENANCE - PAGE 3 BIPAP SYNCHRONY SYSTEM PREVENTIVE MAINTENANCE S CHEDULE (FACTORY RECOMMENDED) Model Number:_______________________________________.
[Page 45] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032412, REV. 00ROUTINE MAINTENANCE - PAGE 4 This page intentionally left blank. .
[Page 46] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TROUBLESHOOTING & DIAGNOSTICS - PAGE 11032443, REV. 00 TROUBLESHOOTING AND DIAGNOSTICS CHAPTER OVERVIEW This section contains information necessary to troubleshoot and diagnose problems with the.
[Page 47] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032443, REV. 00TROUBLESHOOTING & DIAGNOSTICS - PAGE 2 TROUBLESHOOTING FLOW CHART Problem Reported Attempt to Verify Problem by: • Visual Inspection • Operate Unit & Observe Behavior �.
[Page 48] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TROUBLESHOOTING & DIAGNOSTICS - PAGE 31032443, REV. 00 ALARM INTRODUCTION This section describes the BiPAP Synchrony alarms, how to set them, and what corrective actions to take for the alarm co.
[Page 49] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032443, REV. 00TROUBLESHOOTING & DIAGNOSTICS - PAGE 4 OVERVIEW OF ALARM BEHAVIOR Alarm conditions are signalled by the BiPAP Synchrony in three ways: sound, an LED, and a display message. Each.
[Page 50] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TROUBLESHOOTING & DIAGNOSTICS - PAGE 51032443, REV. 00 ALARM LED BEHAVIOR RED ALARM LED The red alarm LED indicates high priority system and patient alarms. The LED flashes when a new high prio.
[Page 51] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032443, REV. 00TROUBLESHOOTING & DIAGNOSTICS - PAGE 6 CARD ERROR ALARM The Card Error alarm is a low priority alarm. It indicates that a problem exists with the SmartCard inserted in the devic.
[Page 52] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TROUBLESHOOTING & DIAGNOSTICS - PAGE 71032443, REV. 00 PATIENT ALARMS The BiPAP Synchrony has the following patient alarms: • Apnea • Patient Disconnect • Low Minute Ventilation • Low Ti.
[Page 53] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032443, REV. 00TROUBLESHOOTING & DIAGNOSTICS - PAGE 8 LOW TIDAL VOLUME ALARM The Low Tidal Volume alarm is a high priority alarm. IT occurs when the calculated tidal volume is less than 90% of.
[Page 54] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TROUBLESHOOTING & DIAGNOSTICS - PAGE 91032443, REV. 00 LOSS OF INPUT POWER This alarm occurs when AC and DC power is lost while the unit is in the Operate state, or AC and DC power is not availa.
[Page 55] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032443, REV. 00TROUBLESHOOTING & DIAGNOSTICS - PAGE 10 TROUBLESHOOTING TABLE Symptom Cause Verification Corrective Action Display Indicator- LED or other indicator not working Main printed c.
[Page 56] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TROUBLESHOOTING & DIAGNOSTICS - PAGE 111032443, REV. 00 Symptom Cause Verification Corrective Action Humidifier will not turn on. Humidifier not turned back on after BiPAP Synchrony was turne.
[Page 57] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032443, REV. 00TROUBLESHOOTING & DIAGNOSTICS - PAGE 12 Symptom Cause Verification Corrective Action No Stored Therapy Data SmartCard not installed Not a Respironics SmartCard (error code d.
[Page 58] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TROUBLESHOOTING & DIAGNOSTICS - PAGE 131032443, REV. 00 Symptom Cause Verification Corrective Action Outlet air temperature- The outlet air temperature is too warm. Filters dirty, blower, Ma.
[Page 59] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032443, REV. 00TROUBLESHOOTING & DIAGNOSTICS - PAGE 14 Symptom Cause Verification Corrective Action Device does not operate from the 12 VDC power source- Faulty DC plug, low DC voltage, fu.
[Page 60] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TROUBLESHOOTING & DIAGNOSTICS - PAGE 151032443, REV. 00 BIPAP SYNCHRONY SYSTEM ERROR CODE RETRIEVAL The BiPAP Synchrony Ventilatory Support System stores error codes that may aid the technician .
[Page 61] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032443, REV. 00TROUBLESHOOTING & DIAGNOSTICS - PAGE 16 SYSTEM ERROR CODES LIST (PATIENT ACTION) Error Code Description of Error Device Action E0 No Error Reserved E1 Generic Software Error REBO.
[Page 62] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TROUBLESHOOTING & DIAGNOSTICS - PAGE 171032443, REV. 00 Error Code Description of Error Device Action E25 Excessive Drift on Flow Sensor STOP E26 Communications Failure with LCD Driver Chip STOP.
[Page 63] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032443, REV. 00TROUBLESHOOTING & DIAGNOSTICS - PAGE 18 Error Code Description of Error Device Action E54 Motor Current High While Blower Off STOP E55 Buffer Overflow REBOOT E56 Motor Current Hi.
[Page 64] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TROUBLESHOOTING & DIAGNOSTICS - PAGE 191032443, REV. 00 NOTE The error conditions are categorized into three groups: CONTINUE/LOGGED: This group contains error conditions in which there is reas.
[Page 65] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032443, REV. 00TROUBLESHOOTING & DIAGNOSTICS - PAGE 20 BIPAP SYNCHRONY SYSTEM ERROR CODES (CORRECTIVE ACTION) Error Code Description of Error Probable Cause Corrective Action E0 No Error Reser.
[Page 66] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TROUBLESHOOTING & DIAGNOSTICS - PAGE 211032443, REV. 00 Error Code Description of Error Probable Cause Corrective Action E23 Invalid Zero Flow Value in Calibration Table Improper Calibration O.
[Page 67] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032443, REV. 00TROUBLESHOOTING & DIAGNOSTICS - PAGE 22 Error Code Description of Error Probable Cause Corrective Action E38 Real-Time Clock not Ticking Defective Microprocessor Defective Cryst.
[Page 68] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TROUBLESHOOTING & DIAGNOSTICS - PAGE 231032443, REV. 00 Error Code Description of Error Probable Cause Corrective Action E55 Buffer Overflow Defective PCA Replace Main PCA E56 Motor Current Hig.
[Page 69] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032443, REV. 00TROUBLESHOOTING & DIAGNOSTICS - PAGE 24 Error Code Description of Error Probable Cause Corrective Action E74 Reserved for Future Use Defective RAM chip Replace Main PCA E75 Inco.
[Page 70] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION TROUBLESHOOTING & DIAGNOSTICS - PAGE 251032443, REV. 00 BIPAP SYNCHRONY SMARTCARD ERROR CODES Error Code Description of Error User Action C1 Unable to Write to SmartCard Card inserted Upside-dow.
[Page 71] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032443, REV. 00TROUBLESHOOTING & DIAGNOSTICS - PAGE 26 This page intentionally left blank. .
[Page 72] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR & REPLACEMENT - PAGE 11032444, REV. 00 REPAIR AND REPLACEMENT CHAPTER OVERVIEW This chapter illustrates the names and locations of the replaceable components in the BiPAP Synchrony system.
[Page 73] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032444, REV. 00REPAIR & REPLACEMENT - PAGE 2 WARNINGS AND CAUTIONS REPLACEMENT INSTRUCTIONS Refer to Figure A and Figure B before removing or installing any components. These figures will det.
[Page 74] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR & REPLACEMENT - PAGE 31032444, REV. 00 F IGURE A: REMOVAL FLOW CHART Alarm Label Key Pad Top Enclosure Main PCA Outlet Port/ Flow Element Diffuser/ Coupler Valve Assembly Blower .
[Page 75] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032444, REV. 00REPAIR & REPLACEMENT - PAGE 4 F IGURE B: INSTALLATION FLOW CHART Bottom Enclosure with foam Buzzer Assembly Muffler Blower Assembly Valve Assembly Diffuser/ Coupler Outlet.
[Page 76] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR & REPLACEMENT - PAGE 51032444, REV. 00 TOP ENCLOSURE REPLACEMENT PROCEDURE Removed/Installed During Process: • Top Enclosure F IGURE C: REMOVAL AND INSTALLATION OF THE TOP ENCLOSURE To .
[Page 77] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032444, REV. 00REPAIR & REPLACEMENT - PAGE 6 F IGURE D: SCREW LOCATIONS 3. While securely holding the top and bottom enclosure together, carefully return the device to its upright position. .
[Page 78] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR & REPLACEMENT - PAGE 71032444, REV. 00 MAIN PRINTED CIRCUIT ASSEMBLY (PCA) REPLACEMENT PROCEDURE Removed/Installed During Process: • Top Enclosure (Refer to page 5 for detailed instruct.
[Page 79] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032444, REV. 00REPAIR & REPLACEMENT - PAGE 8 F IGURE F: LOCATION OF BLOWER CABLE CONNECTOR AND BUZZER CONNECTOR FIGURE G: PRESSURE TUBING FIGURE H: FLOW SENSOR TUBING AND VALVE ASSEMBLY Blower.
[Page 80] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR & REPLACEMENT - PAGE 91032444, REV. 00 7. Remove the Key Pad (Refer to page 10 for more detailed instructions on removing the Key Pad). To install the Main PCA: 1. Install the Key Pad (R.
[Page 81] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032444, REV. 00REPAIR & REPLACEMENT - PAGE 10 KEYPAD REPLACEMENT Procedure Removed/Installed During Process: • Top Enclosure (Refer to page 5 for more detailed instructions.) • Main PCA (R.
[Page 82] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR & REPLACEMENT - PAGE 111032444, REV. 00 To install the Key Pad: 1. Place the key pad on the Main PCA. Align the locking pins with the three holes on the dis- play assembly. Press down o.
[Page 83] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032444, REV. 00REPAIR & REPLACEMENT - PAGE 12 OUTLET-FLOW-TUBING REPLACEMENT Procedure Removed/Installed During Process: • Top Enclosure (Refer to page 5 for more detailed instructions.) •.
[Page 84] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR & REPLACEMENT - PAGE 131032444, REV. 00 To install the Outlet-Flow-Tubing: 1. Position the Outlet Port in its mounting location in the Bottom Enclosure. 2. Connect the Flow Element to the.
[Page 85] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032444, REV. 00REPAIR & REPLACEMENT - PAGE 14 DIFFUSER ASSEMBLY REPLACEMENT Procedure Removed/Installed During Process: • Top Enclosure (Refer to page 5 for more detailed instructions.) • .
[Page 86] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR & REPLACEMENT - PAGE 151032444, REV. 00 VALVE ASSEMBLY REPLACEMENT Procedure Removed/Installed During Process: • Top Enclosure (Refer to page 5 for more detailed instructions.) • Main.
[Page 87] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032444, REV. 00REPAIR & REPLACEMENT - PAGE 16 6. Slide the elbow off of the blower outlet. 7. Lift the Valve Assembly up and away from the bottom enclosure. To install the Valve Assembly: 1. P.
[Page 88] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR & REPLACEMENT - PAGE 171032444, REV. 00 BLOWER ASSEMBLY Procedure Removed/Installed During Process: • Top Enclosure (Refer to page 5 for more detailed instructions.) • Main PCA (Refer.
[Page 89] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032444, REV. 00REPAIR & REPLACEMENT - PAGE 18 To install the Blower Assembly: 1. Place the blower doughnut onto the bottom of the Blower Assembly. Ensure the keys on the blower bottom are al.
[Page 90] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR & REPLACEMENT - PAGE 191032444, REV. 00 MUFFLER REPLACEMENT Procedure Removed/Installed During Process: • Top Enclosure (Refer to page 5 for more detailed instructions.) • Main PCA (R.
[Page 91] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032444, REV. 00REPAIR & REPLACEMENT - PAGE 20 To remove the Muffler: 1. Remove the Top Enclosure (Refer to page 5 for more detailed instructions). 2. Remove the Main PCA (Refer to page 7 for m.
[Page 92] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR & REPLACEMENT - PAGE 211032444, REV. 00 BUZZER ASSEMBLY Procedure Removed/Installed During Process: • Top Enclosure (Refer to page 5 for more detailed instructions.) • Main PCA (Refer.
[Page 93] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032444, REV. 00REPAIR & REPLACEMENT - PAGE 22 BOTTOM ENCLOSURE Procedure Removed/Installed During Process: • Top Enclosure (Refer topage 5 for more detailed instructions.) • Main PCA (Refe.
[Page 94] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR & REPLACEMENT - PAGE 231032444, REV. 00 To install the Bottom Enclosure: 1. Remove the backing from the AVAPS Label. Place the label on the front of the unit, in the bottom right-hand c.
[Page 95] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032444, REV. 00REPAIR & REPLACEMENT - PAGE 24 F IGURE W: FOAM PLACEMENT 7. Install the Buzzer (Refer to page 21 for more detailed instructions). 8. Install the Muffler (Refer topage 19 for mo.
[Page 96] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR KITS - PAGE 11032445, REV. 00 REPAIR KITS CHAPTER OVERVIEW This chapter illustrates the names and components for each of the repair kits for the BiPAP Synchrony system. For technical assista.
[Page 97] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032445, REV. 00REPAIR KITS - PAGE 2 REPAIR KIT REFERENCE TABLE PART NUMER REPAIR KIT NAME PAGE IDENTIFIER 1031394 Top Enclosure Page 3 1031732 Main PCA Page 4 1031731 KeyPad Page 5 1025184 Outlet-.
[Page 98] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR KITS - PAGE 31032445, REV. 00 TOP ENCLOSURE REPAIR KIT PART NUMBER: 1031394 Included in Kit Tools Required Top Enclosure Phillips Screwdriver (No. 2 Medium) .
[Page 99] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION 1032445, REV. 00REPAIR KITS - PAGE 4 MAIN PRINTED CIRCUIT ASSEMBLY (PCA) REPAIR KIT PART NUMBER:1031732 Included in Kit Tools Required Main PCA Phillips Screwdriver (No. 2 Medium) SmartCard Sea.
[Page 100] Respironics BiPAP Synchrony
BIPAP SYNCHRONY SERVICE & TECHNICAL INFORMATION REPAIR KITS - PAGE 51032445, REV. 00 KEYPAD REPAIR KIT PART NUMBER: 1031731 Included in Kit Tools Required KeyPad Phillips Screwdriver (No. 2 Medium) .